Chapter 4. PACS spectrometer scientific capabilities

Table of Contents

4.1. Diffraction Losses
4.2. Grating efficiency
4.3. Spectrometer filters
4.4. Spectrometer relative spectral response function
4.5. Spectrometer field-of-view and spatial resolution
4.6. Spectrometer Point Spread Function (PSF)
4.7. Spectrometer spectral resolution and instrumental profile
4.7.1. Spectrometer spectral resolution
4.7.2. Wavelength calibration
4.7.3. Instrumental profile
4.8. Spectral leakage regions
4.9. Spectrometer flux calibration
4.9.1. Recovering full beam line fluxes and flux densities for point sources
4.9.2. Absolute flux calibration, using the calibration block
4.10. Spectrometer sensitivity
4.11. Spectrometer saturation limits

4.1. Diffraction Losses

The image slicer is the most critical element of the PACS optics, in the figures below the effect of diffraction/vignetting by the entrance field stop and Lyot stop have been included. For the Lyot stop a worst-case loss of 10% is used. For the losses in the spectrometer the fraction of power arriving at the detector is shown in Figure 4.1.

Diffraction throughput of the spectrometer optics; the diffraction losses mainly occur in the image slicer.

Figure 4.1.  Diffraction throughput of the spectrometer optics; the diffraction losses mainly occur in the image slicer.